Measurement Status, potential optics for replacement of H1 core optics.
4ITM05 is the best candidate replacement due to good radius of curvature match with the existing Y-arm ITM
Content moderated by GariLynn Billingsley
billingsley_g  "AT"  ligo.caltech.edu
COC Home page
ALL LIGO 1 OPTICS
4ITM05
Notes
Scatter (small integrating sphere)
Scatter (arb. units, 45 °)
31 ppm
231

Transmission (%)
2.8

AR coating (ppm)
1000
good job Liyuan and Helena!
Bulk Absorption (ppm/cm)
 3 prelim
Data at center only

Coating Absorption ppm (Stanford, witness samples)
HR side absorption (CIT - ppm)
AR side absorption (CIT - ppm)
~0.35
~.5
~.5
 
Radius of Curv. (Km)
measured over 150mm diameter
13.91
See note 1 below
 Image-TPA
 
Surface error, central 80mm
(nm rms)
0 .4406
 
 
 
 
 Notes
 note2 

p - preliminary measurement

Note1:  The range of radii of curvature measured for 4ITM05 were from 13.77K to 14.02K with the Average being 13.91K. The change in the measured saggita over this range is 3.3nm 
The preliminary measurement for this optic 13.58 was measured with a different insrument calibration (the instrument has since been rebuilt)  The new calibration (4-2005) is ... new, and will be refined with ongoing measurement.  The old measurement may be a better indicator of ROC match with other optics measured under the same instrument calibration.